nanoZ User Manual
10
U&9'6+!5:*='5!
When using any of the automated modes of operation, the nanoZ stores
impedance and electroplating results in an Excel-like spreadsheet.
To view or hide the report
window, select View | Report
from the main menu.
The upper panel of the report
shows information about the
currently selected probe and
when it was most recently
tested, user-defined criteria for
flagging whether or not an
electrode site is faulty, and
several display options.
If an electrode has been selected in the main nanoZ program, the probe name
and description is shown. This metadata is extracted from the ʻelectrodes.iniʼ file.
The time and date when the last impedance test or plating procedure began is
also displayed.
The Short and Open user input boxes are editable. An electrode is considered
shorted when the measured impedance magnitude falls below the value entered
in Short, and is considered open circuit when the impedance magnitude is
greater than the value entered in Open. Impedances within this range are
considered normal.
The left-most column in the spreadsheet displays the electrode site index.
The column titles show the impedance test frequency and, if applicable, the
plating current that was applied. In electrode impedance spectroscopy mode
there will be additional columns, two for each test frequency.
The body of the spreadsheet contains the most recent site impedance
measurements for all tested sites. If the Phase checkbox is checked, then both
the impedance magnitude and phase are displayed. If unchecked, only the
impedance magnitude is shown. For electroplating modes, the cumulative plating
time is shown for each individual electrode, which may vary from site to site, for
example, if impedance matching mode was run. Both pre- and post-plating
impedances are stored in separate columns to the left and right of the cumulative
plating time, respectively.
If the Statistics checkbox is checked, then the average impedance magnitude
and phase ± standard deviation of all tested sites is shown at the bottom of each
column. Note that if the Condition checkbox is checked, only ʻnormalʼ sites are
included in these aggregate statistics.