Vue de la page 4
1
-&%.'/&!+'!+0&!*#*'1!
The nanoZ was specifically designed for multichannel microelectrode arrays. It
features extremely low test currents for measuring impedance or electrode
impedance spectroscopy (EIS). Several inbuilt electroplating modes are provided
for automated electrode impedance matching, activation, and cleaning.
!
The setup and operation of the nanoZ is described in this user manual, including
detailed information about the Matlab Software Development Kit (SDK) for
developing your own, customized nanoZ applications.
-0#+R,!:*!+0&!$'DS!
Each nanoZ comes supplied with:
o nanoZ device
o 1.8m USB cable
o NZA-DIP16 adaptor
o NZ-CAL test adaptor
o 3-pin to alligator clip cable
o Installation CD
o (4) stick-on feet (optional)
If any item is missing or appears to be damaged or faulty, please contact the
distributor from whom you purchased your nanoZ.
FEATURES
64 channels
1k ~ 100MΩ working range
1Hz ~ 5kHz test frequencies
1kΩ display resolution, 1% accuracy
or better for 5k ~ 15MΩ electrodes at
test frequencies < 2kHz
low (nanoamp) test currents suitable
for in vitro or in vivo testing
constant current electroplating with
±12μA range, ±5V compliance
intuitive graphical user interface
software development kit for Matlab