nanoZ User Manual
4
To use the nanoZ with in vitro
electrode arrays from MultiChannel
Systems, plug the MCS MEA adaptor
into the nanoZ, then turn the nanoZ
face down and plug it into the MEAʼs
68 pin Honda socket, as shown here.
If the probe is connected directly to
the nanoZ (e.g. a NeuroNexus A32 or
A64 packaged probe), use a short
jumper wire to connect one of the ʻGʼ
pins on the probe PCB to the bath electrode. If using a NZA series adaptor, such
as the supplied NZA-DIP16, connect the bath electrode to the adaptor using the
3-pin-alligator cable supplied. A piece of platinum or platinum-iridium alloy wire
immersed in the bath solution makes an ideal reference connection, as does a
silver-silver chlorided wire or other inert metal. With the MCS MEA adaptor
(shown above) the reference connection of the bath is already connected to the
nanoZʼs reference input, so no external reference wire is required. For other 3
rd
party adaptors, refer to the documentation provided to determine the appropriate
point of connection to the bath electrode.
Take care not to expose the nanoZ to liquids of any kind. If liquid gets
splashed on the nanoZ, wipe it off with a moist cloth and allow to dry. If
solution gets spilled inside the nanoZ, immediately unplug it and remove
any attached electrodes or electrode adaptors. For water spills, allow the nanoZ
to dry completely before resuming use. For saline or other solutions, remove the
two screws that secure the end-cap closest to the probe connector. Slide out the
printed circuit board, being careful not to damage the probe connector on the
enclosure opening. Flush with distilled water to remove all traces of the spillage,
and allow the circuit board to completely dry before re-assembling the nanoZ.
F6'$&!#=#9+'6,!
Probes or electrode arrays that have Samtec connectors (i.e. model number
MOLC11001SQ) can be plugged
directly into the nanoZ, assuming they
conform to the nanoZʼs native pin
mapping (see Appendix C).
The nanoZ comes supplied with a
NZA-DIP16 adaptor for probes that
use a DIP16 connector, and a variety
of other adaptors are available for
commonly used interconnects such as
those from Omnetics, Millmax, and
Hirose. Contact your vendor for more
Omnetics NZ-EIB-36 (Neuralynx)