White Matter NanoZ Manuel d'utilisateur Page 17

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 41
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 16
13
(electrode positive) in ~100nA steps by moving the position of the slider with the
mouse or arrow keys. The voltage across the electrode site will be displayed on
the DMM. The DMM has limited measurement resolution (39mV) and should be
considered approximate, however the calibrated current reading adjacent to the
slider is accurate to within a few nA. The DMM also indicates if the voltage is out
of compliance, that is, if the voltage needed to achieve the desired current is
greater than the nanoZ can generate. To stop applying the current, select Off.
V/9&=#*.&!+ &, +!/'=&!
This mode rapidly cycles through all (or a subset) of the channels on the selected
probe, measuring the impedance of each electrode. The impedance results are
tabulated in the Report window (View | Report).
(A) Set the impedance Test frequency to the desired value. The actual test
frequency that will be used is shown in the status bar.
(B) Choose the number of Test cycles. The default setting of 40 cycles is a good
trade-off between accuracy and speed. Increasing the number of cycles will
marginally improve the accuracy due to averaging, at the expense of a longer
testing duration, whereas decreasing the number of cycles may lower accuracy
due to insufficient localization of the test signal in the frequency domain. The
testing time per site is displayed in the status bar.
(C) The Pause setting specifies a delay, in millisconds, after switching channels
before impedance testing is begun on the currently selected electrode. In most
cases this delay can be set to zero, however with some high capacitance
electrodes the test signal may take some time to settle to baseline after switching
A
B
C
D
Vue de la page 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 40 41

Commentaires sur ces manuels

Pas de commentaire